Abstract
Background: Patients receiving radiotherapy, a clinical treatment process in which radiation is used for the treatment of various types of cancer, utilize a variety of radiation
sources with unique characteristics and procedures. In vivo dose measurements
can help identify systematic
and random
errors in delivery of the
treatment and therefore play an
important role in quality assurance. Recently,
the photon response of optical fibers has been investigated by many research groups. The small diameters of optical
fibers increase the possibility of producing a dosimeter with high spatial resolution, important in the sense that an accurate value
for the absorbed dose in the surrounding
tissue of the dosimeter can be more accurately reported. Another important
advantage of optical fibers as radiation dosimeters is that, unlike conventional TLDs, optical fibers are impervious to water. The aim of this study is to verify
the dosimetric use of fiber
optics in interface dosimetry and to investigate a novel enhanced
dose technique using different
gold thicknesses as a coating
for fiber optics.
Methods: To achieve this goal, commercially available Ge-doped SiO2 optical fibers (Cor Active, Canada) with a core diameter
of 50.9 ± 4.1 µm were irradiated using
a 250 kVp superficial X-ray
machine and a dose of 3 Gy. Before irradiation, fiber optics were prepared
and the following steps, preheat annealing and reading, were performed.
Results: The results show enhancement with increasing gold (Au) thickness, with the highest percentage dose enhancement of
approximately 160% obtained at 80 nm. A slight deviation from the enhancement was obtained at 20 nm, the first thickness
of gold. Encouraging results from such studies
have paved the way for the development of optical fiber
radiation dosimeters
specifically tailored to the task of dosimetry
in radiotherapy.
Conclusion: An optical fiber dosimeter can be placed within the tissue of interest, which is applicable due to its flexibility.
Keywords: Fiber optics, Gold coating, In vivo
dosimetry, Radiotherapy
1. Introduction
As a consequence of
using radiotherapy in approximately 40-60% of all cancer cases1, the accuracy
of this modality is being continuously improved and monitored. Wherefore, all
disciplines of radiotherapy are striving to offer accurate medical care for each
patient. To achieve this goal, the main action will be to improve dosimetric
techniques and protocols2. As a procedural application of the international
recommendations3, in vivo measurements of dose can help identify systematic and
random errors in the delivery of radiation treatment and therefore play an
important role in quality assurance of treatment4. Hence, clinical scientists
have proven that a large variety of radiation dosimetry methods aim to quantify
the amount of energy that is deposited upon interaction with ionizing
radiation2. Among these varieties, one of the most common dosimeters worldwide
is thermoluminescent (TL), which is a characteristic of many crystalline
materials. When such a crystal is exposed to radiation, a very minute fraction of
the absorbed energy is stored in the crystal lattice in electron traps formed
from dopants/ impurities. If the material is subsequently heated, some of this
energy can be recovered as visible light5, the intensity of which is directly
proportional to the original radiation intensity. For many organizations,
thermoluminescence dosimetry (TLD), the most widely used cost-effective
dosimetric technique, has been chosen for routine occupational radiation
exposure monitoring. While the phenomenon of TL can be observed in many
materials, relatively few of these materials have been found to have good
dosimetric characteristics6.
Basically, an optical
fiber is a transparent material used to efficiently transmit light and plays an
important role in daily life. Its technology is widely used in diverse civilian
applications, including communication networks and sensors for the detection of
corrosion in the aircraft industry7. Moreover, different biomedical
applications have been developed, ranging from laser delivery systems, to
disposable blood gas sensors, to intra-aortic probes8. The importance of fiber
optics for radiation dosimetry has been increasingly recognized over the past
decade for dosimetric characterization. The British Research Group has
investigated Ge-doped optical fibers as radiation dosimeters, which provide
good reproducibility and a linear response to doses over a wide dose range,
with a low limit of detectable dose9. Clinical scientists have investigated the
TL performance of irradiated fibers, which have shown considerable potential
for dosimetric applications in radiotherapy, due to their small size, high flexibility,
ease of handling and low cost in addition to their good dose response and low
fading9,10. The TL characteristic of SiO2 is based on trapping processes, most
easily described for crystalline media and is dependent on the presence of
structural defects in the material due to the presence of the doped material10.
The TL response of doped SiO optical fibers has now been investigated using a
wide range of radiation types and energies9,11-14 to investigate photon
response. Other groups have used electron beam irradiation11 and proton beam
irradiation15. Recently, in the UK, in vivo dosimetry (IVD) has been shown to
be an essential procedure for safer radiotherapy16. Hence, studies at the
Egyptian National Cancer Institute, compared different dosimetric tools in the
modality of spatially fractionated radiotherapy (SFRT) and recommended the
establishment of a new dosimetric approach that can detect a small field of 1
cm with high reproducibility and accuracy17,18. Therefore, this study used
commercial telecommunication fiber optics coated with gold to enhance the
dosimetric behavior of patients18, especially for small field size
radiotherapy17.
2. Materials and
MethodsCommercial Ge-doped
SiO2 optical fibers; Cat (MM-50/125): optical properties: losses @850 nm ≤10
dB/ km @ 900 nm & losses @1300 nm ≤10 dB/km, {typical core diameter 50 µm//
cladding diameter 125.0 ±0.1 µm that provided by Corrective, Canada, were used
in this study. To investigate the enhancement achieved by the high atomic
number of the gold coating for the dose measured by the fiber optics, the
following practical methodology was used:
2.1. Preparation of
fiber optics
A. Remove of the acrylate coating: The outer acrylate coating is carefully removed
using a fiber mechanical stripper. The remaining fiber is then cleaned using a
pad of cotton moistened with a small amount of methanol to remove any residual
polymer coating12. (Figure 1)
Figure 1: Shows the
tools and instruments used to prepare fiber optics as a dosimeter;
A: CT 30
cleaver (Fujikura, Japan);
B: vacuum tweezer;
C: storage of the
fiber optics in gelatin capsules retained in an opaque cover;
D: irradiation room,
showing the setup of irradiation; E: TOLEDO TLD readout machine;
F: solar fan motor and
fiber optic; G: coating instrument;
H: the phantom and its
different steps.
B. Cutting the fibers: small pieces of fiber optics were cut and divided into two
groups: one group was coated with a gold layer and the other was the uncoated
control group. A total of 140 pieces of fiber optic tissue, each with a length
of 1 ± 0.1 cm and another group of fibers of the same number, cut into lengths
of
0.5 ± 0.1 cm, were
prepared by a CT 30 cleaver (Fujikura, Japan), the results of which are
presented in Figure 1A. To reduce the uncertainty in TL yield, the gross TL
yield was normalized to the unit mass of each irradiated fiber and an
electronic balance was used to ensure that the mean mass of the fiber and
standard deviation were 4.5 ± 0.02 mg.
C. Storage and handling of the fiber pieces: Vacuum tweezers, which are shown in
Figure 1B, were used to handle the fiber optic pieces19. To ensure unimpeded
reception of luminescence yield by the photomultiplier tube arrangement within
the TLD reader, it is essential that TLD materials do not become contaminated,
particularly with grease19. Moreover, the TL sensitivity, stability,
reusability, precision and minimum detectable dose for optical fibers may be
affected by other physical and environmental factors20. Then, the optical
fibers were placed into gelatin capsules to avoid exposure to elevated humidity
and corrosive agents because these environmental parameters have been shown to
reduce sensitivity by enhancing fading21. In addition, the capsule temperature
decreases when exposed to high ambient light levels, the optical fibers were
kept in a darkened environment prior to being irradiated or read out22. Figure
(1C) shows the fiber optic encapsulation and storage methods.
D. Annealing procedures: The fibers were first annealed in a furnace (Carbolite,
UK) before any irradiation or subsequent TL measurements were taken. For
annealing purposes, the optical fibers were placed in a ceramic boat and
covered with aluminum foil before being placed in the middle of the furnace.
The temperature in the furnace was gradually increased from room temperature to
420°C, after which the mixture remained at that temperature for a period of 1
hour. To minimize thermal stress, the samples were then left in the oven for 18
hours to finally equilibrate at room temperature. This annealing step of TL
material is carried out when the fibers are to be used for the first time or to
be reused for three reasons: to find a good combination of annealing
temperature and time for erasing any effect of previous irradiation, to produce
the lowest intrinsic background and the highest sensitivity and to obtain
reproducibility for both TL and background signals21.
E. Screening process: Prior to the use of TL dosimeters in clinical practice, they should be screened by irradiating them with a known dose from a calibrated radiation source, which is related to the conditions for which the dosimeters will be used22. This process, which is shown in Figure 1D, enables the selection of samples that show good beam uniformity. Figure (1E) shows the instrument used to measure and choose a suitable fiber optic yield. Any TL sample outside the specified tolerance limits was rejected (see Figure 2 A). According to ICRU specifications, this was chosen to be within a limit of ±5% of the group mean3, which is illustrated in Figure 2B, (Figure 2).
Figure 2: A: The TL yield of the optical fibers for each individual (the screening curve).
B: The TL yield
variation for the remaining optical fibers after the selection process.
F. The Glow Curve: The common method of presenting TL data is to plot light
intensity against temperature or time, which is called the glow curve
(represented in Figure 3). With increasing temperature, the light intensity
increases as more electrons are released from the trap and allowed to excite. A
decrease in light intensity occurs as the trap is gradually depleted. The
number of peaks in a glow curve corresponds to the number of different types of
traps existing in the TL material. The temperature at which the maximum of each
peak occurs is correlated with its energy depth, E. The area under a glow curve
is related to the number of electrons trapped and, in turn, corresponds to the
quantity of absorbed radiation23. One of the important experimental problems in
TL is the presence of several overlapping peaks within the TL glow curve. Very
few experimental methods exist that allow decomposition of the TL glow curve
into its individual components24.
G. Gold coating process: First, 1 cm long pieces of fiber optics were prepared for
coating by adhering them to small discs of hard card paper. These discs were
centrally located on a miniature low inertia solar fan motor, the latter being
protected by an aluminum foil cover that also covered the battery. Figure (1F)
shows the low inertia solar fan motor and fiber optic samples attached to the
centrally located card paper disc.
The five discs with
ten fibers per disc were coated with gold in five steps as follows:
- 20 nm thickness - 40
nm thickness - 60 nm thickness - 80 nm thickness - 100 nm thickness. Figure
(1G) shows the coating instrument used (K675X Turbo Large Chromium Coater 8;
Quorum Technologies Ltd., Kent, UK). After the coating, the optical fibers were
ready for irradiation, together with their corresponding uncoated (control)
fibers.
H. Gold etching: The TL yield from the fiber optics is strongly affected by the
presence of gold during reading, so the gold should be removed. The
gold-etching methodology uses aqua regia;
A mixture of
concentrated nitric acid and concentrated hydrochloric acid at a ratio of 1:3
was used for gold removal from the fiber optics25.
2. X-Ray Machines Used in the Experiment
An X-ray machine with
potentials ranging from 150 to 500 kV was used. The operating voltage ranged
between 200 and 300 kV and the current was between 10 and 20 mA. Filters of
different designs are used to achieve half-value layers between 1 and 4mm Cu.
2.1. Irradiation of the fibers: The fibers were put in gelatin capsules and
irradiated at Royal Surrey County Hospital, Guildford, UK, using an
orthovoltage X-ray unit (GULMAY MEDICAL, Surrey UK), which has a 1 mm lead
window. The half-value layer HVL is 2.7 mm Cu, the FSD is 50 cm, the added
filtration in mm is 1.5 mm Al + 0.25 mm Cu + 0.5 mm Sn, the tube current is 12
mA and the tube potential is 250 kVp. The optical fibers were exposed to a 3 Gy
single dose.
2.2. Reading the fibers for screening: The two requirements for determining the dose
given to the fiber optics are a reliable form of heating and a method of
measuring the light output19. To maintain control of thermal fading, the
readout was performed after a set period post irradiation of 12 hours. The TLD
reader used was a TOLEDO system (Pitman Instruments, Weybridge, UK), which is
presented in Figure 1E.
The readout was
carried out in the presence of nitrogen gas to prevent the influence of
triboluminescence. In addition, this reduces the amount of oxidation that would
otherwise occur on the surface of the dosimeter. Triboluminescence is caused by
the mechanical disturbance of the surface of the fiber when the fiber is cut
into small pieces20.
The parameters that
provided an optimal glow curve and were used during the readout process were as
follows: preheating temperature of 160°C for 10 seconds and readout temperature
of 300°C for 25 seconds with a ramp rate of 25°C/sec. An annealing temperature
of 300°C for 10 seconds was subsequently used to eliminate any residual signal;
its reproducibility was (±1.5) and a low residual signal was obtained for a
readout temperature of 300°C and negligible fading. After reading and
performing the statistical analysis for the screening process, the steps for
coating and subsequent irradiation were carried out.
2.3. The irradiation setup of the fiber optics inserted in the phantom simulated the RT dosimeter:
The optical fibers
were distributed in the phantom in steps according to their gold coating
thickness. The phantom, a model for the synovial membrane, is formed of a
Perspex cube with a length of 7 cm26. The phantom was designed at the
University of Surrey. Figure (1 H) illustrates the shape of this phantom and
its different steps. The irradiation setup was the same as that for the
screening process, with the same monitoring units, time and dose rate.
3. Results and Discussion
3.1. Screening optical fibers
The Ge dopant
distribution in the longitudinal direction of fiber optics is heterogeneous and
provides for TL radiation dosimetry applications; thus, the TL yield will be
affected27.
The selection of fiber
dosimeters with uniform sensitivity is important for allowing further
investigation of the enhancement due to gold coating. Dosimeters were selected
in an effort to eliminate outlier data while retaining optical fibers within ±1
standard deviation of the mean value for subsequent use10. This process of
selecting Ge-doped SiO2 optical fibers was performed by irradiating large
sample groups at a fixed dose as well as at a fixed dose rate (see Figure 2 A).
After the screening
process, the selected fibers were found to be within ± 4.82% of the
standard deviation (SD) of the mean TL value and were selected to minimize the
variation in the TL yield of the fibers in subsequent measurements. As a
result, from the initial 140 samples, only 105 samples were selected,
reflecting that these had dopant concentrations within ± 5% variation (Figure
2.B). For fibers to be used in radiotherapy, the combined uncertainty should be
no more than 5% in well- controlled radiotherapy, which is approximately the
same order as the lowest dose difference that can be detected clinically27.
The coating,
post-coating irradiation and reading of the fiber optics were repeated a number
of times to overcome the difficulties that were found in the gold etching. The
aqua regia was used for gold etching. Unfortunately, there was a visible effect
on the surface of the cleaned fiber (namely, a degree of opacity) due to the
strong etching agent.
3.2. The glow curve
The sensitivity of a
particular TLD material is defined as the TL signal (the peak height or TL
intensity integrated over a certain temperature region) per unit absorbed dose
and per unit mass. This sensitivity depends on the readout system, the heating rate
and the light detection system used6. The effect of heating the fiber optics to
stimulate the release of trapped energy after irradiation can be represented
graphically as the total number of counts recorded against the temperature;
this phenomenon is called the glow curve. The area under the curve represents
the radiation energy deposited into the fiber. (Figure 3) shows the glow curve
of a piece of optical fiber, which was used as a control non-coated fiber. The
curve shows a single peak, as expected for the amorphous material, unlike the
crystals of the TLD material, which show multiple peaks due to localized defect
trapping.
Figure 3: Glow curve
of one of the optical fibers after irradiation with a 3Gy photon
dose.
3.3. The TL yield of the optical fibers
The irradiation setup
was previously described, with the TL yield represented as counts per second
per unit mass of the pieces of the fibers. The resultant relation between the
TL yield and the depth of the fiber in the phantom is graphically represented
in (Figure 4).
(Figure 5) shows the
reproducibility of the results obtained by repeating the experiment. The
following curve shows the results for a different set of optical fibers from
the same manufacturer (CorActive, Canada) with the same diameter and dopant
material. As shown in Figure 4.5, the sensitivity of these new optic fibers
differs from that of the first fibers. Due to this variation, the whole
selection process, including capturing the individual glow curves and recording
the TL yield, was repeated.
Figure 4: Relationship
between the depth at which non-coated optical fibers are
located in the phantom and the TL yield.
Figure 5: Illustrates
the relationship between the distance at which non-coated optical fibers in the
phantom were arranged and the TL yield.
As expected, the curve
shows an approximate exponential dependency on depth in the phantom.
3.4. The enhancement of the TL yield of the gold-coated optical fibers:
(Figure 6) shows the
TL yields of the coated and non-coated fiber optics and a comparison was made
between them. At a gold coating thickness of 80 nm, the highest TL yield was
reported. The lowest value was obtained for a gold thickness of 100 nm.
Figure 6: Shows the
comparison between the TL yield of the gold-coated and
control non-coated fiber optics.
The dose enhancement
factor (DEF) was calculated using the TL yield of the gold-coated fiber optics
and the corresponding control non-coated fibers in the same phantom step, i.e.,
at the same depth inside the phantom and at the same distance from the X-ray
applicator. Equation (1) is used to calculate the dose enhancement factor:
DEF = (TL yield of
gold-coated/TL reading of non-coated fiber) × 100 (1),
(Figure 7)
Figure 6 clarifies that there is evidence of an increase in the dose due to the presence of gold coating the optical fibers. Figure 7 shows two exceptions to the coated fiber enhancement, which occur for 20 nm and 100 nm thick gold. The first thickness of gold may not be sufficient to observe any statistically significant DEF. In contrast, for the 100 nm thick sample, the inefficient removal of residual gold particles by chemical processes is a potential problem, leading to an inhibited TL yield. The enhanced TL yields obtained indicated that a large DEF may be obtained, for instance, at the interface between the synovial membrane surface and gold-coated contrast agents such as microspheres.
Figure 7: Shows the
percentage enhancement in the TL yield due to the presence of gold.
Monte Carlo simulation
(MCNPXv.2.6) was used to confirm the results of the experiment. There was an
increase in the TL yield due to photoelectron production for the two phantoms
used to resemble the inflamed joint in the case of iodine compared with the yield
in the absence of iodine26. The importance of using optical fibers as radiation
dosimeters has recently increased due to their higher chemical stability and
relatively lower effective atomic number than those of films. Moreover, the
addition of impurities or dopants to SiO2 optical fibers enhances the radiation
detection sensitivity28. In contrast to film dosimeters, optical fibers have a
wide dynamic range. Fiber has been investigated by using different types of
radiation, a range of energies and even different techniques. The performance
of optical fibers has shown considerable potential for dosimetry applications
in radiotherapy for both external beam and internal (brachytherapy). Concerning
external radiotherapy, many workers have investigated fiber optics by using
photon and electron beams14 and for brachytherapy29. Moreover, the
reproducibility and accuracy of this improved dose metric behavior need further
study to be applicable in advanced and sophisticated modalities in
radiotherapy, e.g., spatially fractionated radiotherapy (SFRT).
4. Conclusion
This in vivo dosimetry
study at the micro scale was performed using fiber optics, which offers the
advantage of high spatial resolution and low cost because this medium is
reusable and reproducible. Dose enhancement can be used in radiotherapy to
ensure that there is a high uniform dose to the tumor with a minimum dose to
the normal surrounding tissue. The use of gold (Au) in radiation in vivo
dosimetry is promising because it is a high-atomic-number material that is
nontoxic and hence tolerable in the body. The difficulty of etching Au from the
surface of fiber optics is only a disadvantage of this technique. Moreover, the
results show enhancements with thicknesses of 40 nm, 60 nm and 80 nm, with the
highest percentage of the dose enhancement occurring at an 80 nm thickness of
gold, at 159%. There was a slight deviation from the enhancement at the first
thickness of 20 nm gold, which may be due to the insufficient thickness for
collecting the photoelectron products inside the micro dosimeter. In contrast,
at a 100 nm thickness, residual gold particles almost certainly prevent the
detection of the TL yield. The improved dosimetry technique used in this study
could be applied in sophisticated radiotherapy modalities, such as spatially
fractionated radiotherapy (SFRT).
Conflicts of
Interest
The authors declare
that there is no conflict of interest.
Financial Support and
Funding
There is No Funding.
Declarations
Ethics approval and
consent to participate: Not applicable.
Consent for
publication: Authors provide consent of the manuscript to be published.
Availability of data
and material: Data available if requested.
Competing interests:
The authors declare that there is no conflict of
interest.
Funding: The authors
declare that “there is No Fund”.
Authors’ contributions: All authors are
equally contributed.
Acknowledgements: Not applicable
List of
abbreviations
Not applicable